FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4111421
·
Received September 23, 2014
Report
- Report Number
- 2032227-2014-27621
- Event Type
- Malfunction
- Date Received
- September 23, 2014
- Date of Event
- August 25, 2014
- Report Date
- August 25, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
INSULIN PUMP HAD AN INTERMITTENT BUTTON RESPONSE DUE TO CORRODED ACT AND DOWN BUTTON ON KEYPAD TRACE. NO BUTTON ERROR ALARM DURING TESTING. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER AND CRACKED RESERVOIR TUBE LIP.
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER WAS CHANGING THE BATTERY. BLOOD GLUCOSE VALUE IS 206 MG/DL. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 589669 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LNAP | A000114634 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |