FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4111417 · Received September 23, 2014

Report

Report Number
2032227-2014-27614
Event Type
Malfunction
Date Received
September 23, 2014
Date of Event
August 25, 2014
Report Date
August 25, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED DISPLAY WINDOW AND CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS WERE NOTED DURING THE VISUAL INSPECTION. NO BUTTON ERROR ALARMS WERE NOTED. ALL OF THE BUTTONS FUNCTIONED PROPERLY. HOWEVER, MOISTURE DAMAGE WAS NOTED ON THE KEYPAD TRACES.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP IS ALARMED BUTTON ERROR. CUSTOMER DID NOT KNOW HER BLOOD GLUCOSE VALUE. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
590479 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAL A000289606

Patients

Seq Age Sex Outcome Treatment
1 42 YR