FDA Adverse Event Malfunction Summary report: N

OPTICROSS?

MDR report key: 4111341 · Received September 23, 2014

Report

Report Number
2134265-2014-05634
Event Type
Malfunction
Date Received
September 23, 2014
Date of Event
July 14, 2014
Report Date
August 29, 2014
Manufacturer
BOSTON SCIENTIFIC - FREMONT (SUD)
Product Code
OBJ
PMA / PMN Number
K123621
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT THE TIME OF EVENT: 18 YEARS OR OLDER. DEVICE EVALUATED BY MANUFACTURER: THE COMPLAINT DEVICE WAS RECEIVED FOR EVALUATION. EVALUATION OF THE RETURNED DEVICE REVEALED THAT THE TELESCOPE ASSEMBLY WAS NOT ABLE TO PROPERLY PULL BACK, ADVANCE AND RETRACT. THE TELESCOPE CANNOT ADVANCE THE TRANSDUCER DISTAL HOUSING (TDH) TO THE MOST DISTAL POSITION. AN OPEN HOLE WAS OBSERVED AT THE SHEATH LAP JOINT SECTION OF THE DEVICE. FLUID WAS LEAKING FROM THE OPEN HOLE AT THE SHEATH LAP JOINT ASSEMBLY WHEN THE CATHETER WAS FLUSHED. DURING IMAGE CHARACTERIZATION TESTING, NO IMAGE APPEARED IN THE SYSTEM DUE TO ELECTRICAL OPEN AT PROXIMAL. NO OTHER ISSUES OR DEFECTS WERE OBSERVED DURING PRODUCT ANALYSIS OF THE RETURNED DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE OF THE REPORTED DIFFICULTY IS A DESIGN CONSTRAINT OF THE PRODUCT. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON THE DEVICE ANALYSIS COMPLETED ON (B)(4) 2014. IT WAS REPORTED THAT A LOST IMAGE OCCURRED DURING THE PROCEDURE. THE 90% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED OF PROXIMAL TO DISTAL RIGHT CORONARY ARTERY. DURING THE PERCUTANEOUS TRANSLUMINAL CORONARY ROTATIONAL ATHERECTOMY (PTCRA), AN OPTICROSS¿ IMAGING CATHETER WAS USED HOWEVER IMAGE WAS LOST WHEN INSERTED IN THE SEVERELY CALCIFIED AREA. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CONDITION IS GOOD. HOWEVER, DEVICE ANALYSIS REVEALED AN OPEN HOLE AT THE SHEATH LAP JOINT SECTION OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
590808 OPTICROSS? CATHETER, ULTRASOUND, INTRAVASCULAR OBJ BOSTON SCIENTIFIC - FREMONT (SUD) H749518080 17022195

Patients

Seq Age Sex Outcome Treatment
1