MALLEABLE SUCTION MEDIUM, STANDARD TIP
Report
- Report Number
- 1723170-2014-01013
- Event Type
- Malfunction
- Date Received
- September 23, 2014
- Date of Event
- August 25, 2013
- Report Date
- August 25, 2013
- Manufacturer
- MEDTRONIC NAVIGATION, INC.
- Product Code
- PGW
- PMA / PMN Number
- K133665
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
RMA ISSUED. REPLACEMENT MALLEABLE SUCTION SHIPPED TO SITE (B)(4) 2014. MEDTRONIC INVESTIGATION OF RETURNED SUSPECT MALLEABLE SUCTION FINDS THAT THE TOOL SHOWS RED STATUS WITH ERROR "POOR SIGNAL" IN THE SYSTEM ACCURACY TEST. THE SUCTION TOOL WILL NOT NAVIGATE. DIAGNOSTICS IN THE CRANIAL APPLICATION SHOW ONLY 2 OF 3 COILS ARE SENDING DATA. ELECTRICAL FAILURE - DIRECTLY CAUSED EVENT.
A MEDTRONIC ENT REPRESENTATIVE REPORTED THAT, WHILE IN A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS), A MALLEABLE SUCTION UNEXPECTEDLY STOPPED TRACKING. THIS OCCURRED AT APPROXIMATELY 25 MINUTES INTO THE PROCEDURE, PREVIOUSLY THE INSTRUMENT WORKED PROPERLY. TROUBLE-SHOOTING DID NOT RESOLVE THE ISSUE. THE SURGEON OPTED TO USE AN ALTERNATE INSTRUMENT, A FRAZIER STRAIGHT SUCTION, TO CONTINUE AND COMPLETED THE PROCEDURE WITH THE USE OF THE NAVIGATION SYSTEM. THERE WAS NO IMPACT ON PATIENT OUTCOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 590132 | MALLEABLE SUCTION MEDIUM, STANDARD TIP | EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT | PGW | MEDTRONIC NAVIGATION, INC. | 140530J |