FDA Adverse Event Malfunction Summary report: N

MALLEABLE SUCTION MEDIUM, STANDARD TIP

MDR report key: 4111236 · Received September 23, 2014

Report

Report Number
1723170-2014-01013
Event Type
Malfunction
Date Received
September 23, 2014
Date of Event
August 25, 2013
Report Date
August 25, 2013
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
PGW
PMA / PMN Number
K133665
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

RMA ISSUED. REPLACEMENT MALLEABLE SUCTION SHIPPED TO SITE (B)(4) 2014. MEDTRONIC INVESTIGATION OF RETURNED SUSPECT MALLEABLE SUCTION FINDS THAT THE TOOL SHOWS RED STATUS WITH ERROR "POOR SIGNAL" IN THE SYSTEM ACCURACY TEST. THE SUCTION TOOL WILL NOT NAVIGATE. DIAGNOSTICS IN THE CRANIAL APPLICATION SHOW ONLY 2 OF 3 COILS ARE SENDING DATA. ELECTRICAL FAILURE - DIRECTLY CAUSED EVENT.

Description of Event or Problem · 1

A MEDTRONIC ENT REPRESENTATIVE REPORTED THAT, WHILE IN A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS), A MALLEABLE SUCTION UNEXPECTEDLY STOPPED TRACKING. THIS OCCURRED AT APPROXIMATELY 25 MINUTES INTO THE PROCEDURE, PREVIOUSLY THE INSTRUMENT WORKED PROPERLY. TROUBLE-SHOOTING DID NOT RESOLVE THE ISSUE. THE SURGEON OPTED TO USE AN ALTERNATE INSTRUMENT, A FRAZIER STRAIGHT SUCTION, TO CONTINUE AND COMPLETED THE PROCEDURE WITH THE USE OF THE NAVIGATION SYSTEM. THERE WAS NO IMPACT ON PATIENT OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
590132 MALLEABLE SUCTION MEDIUM, STANDARD TIP EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT PGW MEDTRONIC NAVIGATION, INC. 140530J