FDA Adverse Event Malfunction Summary report: N

INDIGO¿

MDR report key: 4111233 · Received September 23, 2014

Report

Report Number
1045254-2014-00231
Event Type
Malfunction
Date Received
September 23, 2014
Report Date
September 1, 2014
Manufacturer
MEDTRONIC XOMED INC.
Product Code
HBE
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE ATTACHMENT WAS EVALUATED AND FOUND TO HAVE CORRODED BEARINGS.

Additional Manufacturer Narrative · 1

(B)(4): THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. METHOD ¿ NO TESTING METHODS PERFORMED FDA. RESULTS ¿ NO RESULTS AVAILABLE SINCE NO EVALUATION PERFORMED. CONCLUSION ¿ DEVICE NOT RETURNED. (B)(4)

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE QUALITY ENGINEERING TEAM. AS RECEIVED CONDITION: RECEIVED INDIGO MOTOR AND ANGLED ATTACHMENT FOR EVALUATION BY THE ENGINEERING GROUP. THE CONDITION OF THE DEVICES SHOWED NO SIGNIFICANT PHYSICAL DAMAGES. ANALYSIS RESULTS: EQUIPMENT USED: TEST IPC CONSOLE (PROVIDED BY S<(>&<)>R). OBSERVATIONS: THE INDIGO MOTOR WAS CONNECTED TO THE IPC AND RUN AT DEFAULT SPEED WITH THE ANGLED ATTACHMENT. THE HANDPIECE FUNCTIONED AND SOUNDED NORMAL. THE DEVICE PASSED TEMPERATURE TESTING WITH THE MEASURED TEMPERATURE AFTER 1 MINUTE OF RUN-IN AT 92.3F. BASED ON THE ABOVE OBSERVATIONS THE UNDERLYING CAUSE COULD BE DETERMINED AS THE REPORTED COMPLAINT ¿HEAT¿ COULD NOT BE CONFIRMED. THE DEVICE WAS REPAIRED AND TESTED TO PRODUCT SPECIFICATIONS AND RETURNED TO THE CUSTOMER. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT ¿DURING THE SURGICAL PROCEDURE THE HANDPIECE HEATS.¿ NO EVENT DATE WAS PROVIDED. THERE WAS NO PATIENT IMPACT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
590138 INDIGO¿ DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE MEDTRONIC XOMED INC. 1845000 206258008

Patients

Seq Age Sex Outcome Treatment
1