FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK CR(R) PLUS DEFIBRILLATOR
MDR report key: 4111202
·
Received September 23, 2014
Report
- Report Number
- 3015876-2014-01114
- Event Type
- Malfunction
- Date Received
- September 23, 2014
- Date of Event
- August 29, 2014
- Report Date
- August 29, 2014
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K033275
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL ADVISED THE CUSTOMER THAT THEIR DEVICE IS NOT FIELD SERVICEABLE AND NO LONGER UNDER WARRANTY. PHYSIO RECOMMENDED THAT THE CUSTOMER PURCHASE A REPLACEMENT DEVICE. THE DEVICE HAS NOT BEEN RETURNED TO PHYSIO-CONTROL FOR EVALUATION. THE CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED.
Description of Event or Problem · 1
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE HAD ALL THREE ICONS (CHARGE PAK, ATTENTION AND SERVICE WRENCH) PRESENT ON THE DISPLAY. THE THREE ICONS BEING ILLUMINATED INDICATES THAT THE DEVICE MAY NOT BE ABLE TO PROVIDE SUFFICIENT DEFIBRILLATION THERAPY, IF NECESSARY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 590523 | LIFEPAK CR(R) PLUS DEFIBRILLATOR | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC | CRPLUS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|