FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4111182 · Received September 23, 2014

Report

Report Number
2032227-2014-22979
Event Type
Malfunction
Date Received
September 23, 2014
Date of Event
April 30, 2014
Report Date
April 30, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UNIT RECEIVED WITH HIGH STOP CURRENT DUE TO FAULTY C4 AT RF BOARD. UNIT RECEIVED WITH CRACKED RESERVOIR TUBE LIP. UNIT RECEIVED WITH MINOR SCRATCHED LCD WINDOW.(B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP'S BATTERY LIFE WAS ONLY ABOUT ONE AND A HALF DAYS. BLOOD GLUCOSE LEVEL AT THE TIME OF THE CALL WAS NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
589685 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-554WWL

Patients

Seq Age Sex Outcome Treatment
1