FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4110970 · Received September 23, 2014

Report

Report Number
2531779-2014-27140
Event Type
Malfunction
Date Received
September 23, 2014
Report Date
September 11, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS OBSERVED TO BE PEELING, WORN AND THINNING. THE UP AND DOWN BUTTONS WERE INTERMITTENTLY UNRESPONSIVE TO TESTING. THE DISPLAY WAS OBSERVED TO BE DIM, FADING, AND REDDISH IN COLOR. THE BOLUS BUTTON COVER WAS OBSERVED TO BE DETACHED. THE KEYPAD COVER WAS REMOVED FOR FURTHER INVESTIGATION AND CONTAMINATION WAS FOUND UNDER ALL CONTACTS.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DIM, RED DISPLAY, INTERMITTENT FUNCTION TO THE UP AND DOWN BUTTONS AND CONTAMINATION ON ALL OF THE BUTTON CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
591376 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 14 YR