FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE(ITEM #UNK)

MDR report key: 4110880 · Received September 16, 2014

Report

Report Number
1419937-2014-00699
Event Type
Malfunction
Date Received
September 16, 2014
Date of Event
September 10, 2014
Report Date
September 10, 2014
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT POWER SUPPLY WAS SENT TO THE CUSTOMER. THE PRODUCT INVOLVED IN THE COMPLAINT WAS NOT RETURNED FOR EVALUATION/INVESTIGATION AT THIS TIME. THEREFORE, NO CONCLUSIONS CAN BE MADE AS TO THE CAUSE OF THE EVENT. SHOULD ADDITIONAL INFORMATION OR THE ORIGINAL PRODUCT BE RECEIVED, RESULTING IN NEW, CHANGED, OR CORRECTED INFORMATION, A FOLLOW UP REPORT WILL BE FILED AT THAT TIME. THIS ISSUE WITH A DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT SHE DROPPED HER AC ADAPTOR AND CRACKED THE HOUSING AND WIRES ARE EXPOSED, WHICH IS A SAFETY RISK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
573841 PUMP IN STYLE(ITEM #UNK) HGX MEDELA, INC. 57XXX/9207010 REV L-1712

Patients

Seq Age Sex Outcome Treatment
1