FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES
MDR report key: 4110805
·
Received September 23, 2014
Report
- Report Number
- 3015876-2014-01112
- Event Type
- Malfunction
- Date Received
- September 23, 2014
- Date of Event
- July 9, 2014
- Report Date
- August 26, 2014
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K102972
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING AND THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED POWER PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A SHORTED TANTALUM CAPACITOR, DESIGNATOR C4.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S DEVICE WOULD NO LONGER POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 589944 | LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |