RESERVOIR 1.8ML
Report
- Report Number
- 2032227-2014-27934
- Event Type
- Malfunction
- Date Received
- September 23, 2014
- Date of Event
- August 24, 2014
- Report Date
- August 24, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- FRN
- PMA / PMN Number
- K001828
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP WAS ALARMING NO DELIVERY. CUSTOMER STATES THAT THEY HAVE CHANGED THE RESERVOIR AND INFUSION SET, YET ISSUE HAS CONTINUED. DURING TROUBLESHOOTING, CUSTOMER WAS ASKED TO PERFORM A FIXED PRIME OF INSULIN. CUSTOMER STATES THE INSULIN DID EXIT. NEXT, CUSTOMER WAS ASKED TO MANUAL PRIME OF INSULIN. CUSTOMER STATES THE INSULIN DID NOT EXIT AND EXPERIENCED GREATER THAN NORMAL RESISTANCE. ADVISED CUSTOMER TO CHANGE RESERVOIR AND SET. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 122 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 589563 | RESERVOIR 1.8ML | RESERVOIR | FRN | MEDTRONIC MINIMED | MMT-326A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |