FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4110772 · Received September 23, 2014

Report

Report Number
2531779-2014-27151
Event Type
Malfunction
Date Received
September 23, 2014
Report Date
September 10, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/03/2014 WITH THE FOLLOWING FINDINGS: THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED AT THE OPENING, DOWN TO RUBBER BUMPER. UNRELATED TO THE COMPLAINT, THE TEXT ON THE DISPLAY SCREEN WAS FADED AND HAD A REDDISH TINT.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE. THE BATTERY COMPARTMENT REPORTEDLY WAS CRACKED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
590366 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1