FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 4110596 · Received September 16, 2014

Report

Report Number
1720753-2014-07925
Event Type
Malfunction
Date Received
September 16, 2014
Date of Event
August 28, 2014
Report Date
September 16, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE HIGH VOLTAGE CABLE AND THE LEMO CONNECTOR WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM PRODUCED UNCOMMANDED X-RAYS DURING A PROCEDURE WITH PT INVOLVEMENT, THEREFORE THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ARO. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
572243 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1