PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-27425
- Event Type
- Injury
- Date Received
- September 23, 2014
- Date of Event
- August 25, 2014
- Report Date
- August 25, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS
- Reporter Occupation
- OTHER
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP HAD NORMAL OPERATING CURRENTS AND NO UNEXPECTED OFF NO POWER OR LOW BATTERY ALARM WAS NOTED. THE INSULIN PUMP PASSED THE FUNCTIONAL TESTING, INCLUDING THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, NO DELIVERY AND MOTOR TESTS. NO MOTOR ERROR ALARM OR MOTOR ANOMALY WAS NOTED. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED RESERVOIR TUBE LIP.
CUSTOMER'S FATHER REPORTED THAT HIS SON'S BLOOD GLUCOSE WAS OVER 400 MG/DL. THE FATHER STATED THAT HE HAD ALREADY CHANGED THE INFUSION SET AND THE INSULIN PUMP WAS NOT DELIVERING INSULIN TO HIS SON. THE SON'S BLOOD GLUCOSE WAS LOWERED WITH A MANUAL INJECTION. THE FATHER STATED THE INSULIN PUMP'S DRIVE SUPPORT CAP APPEARED TO BE NORMAL. IT WAS ALSO FOUND THAT THERE WAS NO AIR IN THE INSULIN PUMP'S TUBING. THE FATHER STATED THAT THERE WAS A PROBLEM WITH THE MOTOR DURING A REWIND OF THE INSULIN PUMP. THE CUSTOMER WAS ADVISED THE INSULIN PUMP NEEDED TO BE REPLACED. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 589959 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-523NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 6 YR |