REMB ELECTRIC WIREDRIVER
Report
- Report Number
- 0001811755-2014-03338
- Event Type
- Malfunction
- Date Received
- September 23, 2014
- Date of Event
- August 26, 2014
- Report Date
- August 27, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K112593
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.
THE REPORTED EVENT WAS NOT DUPLICATED. UPON DISASSEMBLY FOR VISUAL INSPECTION, SOCKETS WERE CORRODED, WHICH MAY CAUSE OR CONTRIBUTE TO THE REPORTED EVENT.
IT WAS REPORTED THAT DURING TESTING AT THE USER FACILITY, THE REMB ELECTRIC WIREDRIVER CONTINUED TO REMAIN ACTIVE UNTIL THE DEVICE WAS UNPLUGGED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
IT WAS REPORTED THAT DURING TESTING AT THE USER FACILITY, THE REMB ELECTRIC WIREDRIVER CONTINUED TO REMAIN ACTIVE UNTIL THE DEVICE WAS UNPLUGGED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.