PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE
Report
- Report Number
- 1419937-2014-00674
- Event Type
- Malfunction
- Date Received
- September 10, 2014
- Date of Event
- August 21, 2014
- Report Date
- August 21, 2014
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
A REPLACEMENT POWER ADAPTER WAS SENT TO THE CUSTOMER AND REQUESTED THE DEFECTIVE POWER ADAPTER BE RETURNED FOR EVAL. THE DEFECTIVE POWER ADAPTER WAS RECEIVED AT MEDELA ON (B)(4) 2014 AND EVALUATED (B)(4) 2014. A PRODUCT EVAL WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED PRONG ISSUE BUT HAD REVEALED THAT THE TRANSFORMER HOUSING WAS BREACHED, EXPOSING INNER ELECTRICAL CIRCUITRY, WHICH IS A SAFETY RISK. THIS ISSUE WITH A DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING. EVAL SUMMARY: A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED THE TRANSFORMER HOUSING WAS BREACHED, EXPOSING INNER ELECTRICAL CIRCUITRY. A VISUAL EXAMINATION OF THE CORD REVEALED NOTHING UNUSUAL. THE POWER SUPPLY WAS PLUGGED IN AND THE VOLTAGE ACROSS THE DC PLUG WAS MEASURED AT 13.9 VDC, WHICH IS NORMAL AND INDICATES THAT THE POWER SUPPLY IS PRODUCING THE CORRECT VOLTAGE. RESISTANCE ACROSS THE DC PLUG WAS MEASURED OPEN, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. SMOKE, FIRE AND SPARKING WERE NOT OBSERVED WHILE THE POWER SUPPLY WAS PLUGGED IN. THE RESISTANCE ACROSS THE AC BLADES MEASURED AS 57.0 OHMS, WHICH IS NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW.
ON (B)(6) 2014, THE CUSTOMER REPORTED THE POWER ADAPTER PRONGS WERE FALLING OUT OF THE TRANSFORMER HOUSING. DURING THE PRODUCT EVAL ON (B)(6) 2014, THE TECHNICIAN OBSERVED A BREACHED TRANSFORMER HOUSING EXPOSING INNER ELECTRICAL CIRCUITRY, WHICH IS A SAFETY RISK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 559908 | PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE | HGX | MEDELA, INC. | 57063/9207010 | 3312 REV M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |