INRATIO PT/INR TEST STRIPS
Report
- Report Number
- 2027969-2014-00839
- Event Type
- Malfunction
- Date Received
- September 12, 2014
- Date of Event
- August 26, 2014
- Report Date
- August 27, 2014
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K092987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
IT IS INDICATED THAT PRODUCT IS NOT RETURNING FOR EVAL. SINCE THE PRODUCT IS NOT RETURNING FOR EVAL. SINCE THE PRODUCT ASSOCIATED WITH THE COMPLAINT WAS NOT RETURNED, A REVIEW OF IN-HOUSE TESTING DATA WAS PERFORMED. RETAIN STRIP TESTING RESULTS MET BOTH ACCURACY AND REPEATABILITY CRITERIA. THE PRODUCT PERFORMED AS EXPECTED AND NO PRODUCT DEFICIENCIES WERE OBSERVED. A REVIEW OF THE MFR RECORD FOR THE LOT DID NOT UNCOVER ANY RELEVANT NON-CONFORMANCE. LOT MEETS RELEASE SPECIFICATION. ROOT CAUSE COULD NOT BE DETERMINED FROM THE INFO PROVIDED BY THE CUSTOMER. BASED ON THE INFO AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
CALLER ALLEGED DISCREPANT INRATIO RESULTS. RESULTS AS FOLLOWS: DATE: (B)(6). INRATIO: 3.8. A 1ST RE-TEST: GREATER THAN 7.5. A 2ND RE-TEST: 5.6. DATE: (B)(6). INRATIO: 2.7. A 1ST RE-TEST: 5.5. A 2ND RE-TEST: GREATER THAN 7.5. MD LAB: 2.7. TIME BETWEEN TESTS: INRATIO TESTS ON (B)(6) WERE ALL PERFORMED WITHIN 9 MINUTES. A 16 HOURS BETWEEN INRATIO TESTS ON (B)(6) AND INRATIO 2.7 INR ON (B)(6). A 6.5 HOURS BETWEEN INRATIO 2.7 INR ON (B)(6) AND MD LAB. LESS THAN 1 HOUR BETWEEN MD LAB ON (B)(6) AND INRATIO TESTS (5.5 AND GREATER THAN 7.5). LESS THAN 5 MINUTES BETWEEN THE INRATIO TESTS THAT WERE PERFORMED AFTER THE MD LAB DRAW ON (B)(6) (5.5 AND GREATER THAN 7.5 INR). THERAPEUTIC RANGE: 2.0-3.0.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 566659 | INRATIO PT/INR TEST STRIPS | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100071 | 335798 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INRATIO MONITOR SERIAL NUMBER: (B)(4)| WARFARIN |