FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 4110257 · Received September 10, 2014

Report

Report Number
2518422-2014-01563
Event Type
Malfunction
Date Received
September 10, 2014
Date of Event
August 18, 2014
Report Date
August 18, 2014
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "VENTILATOR INOPERATIVE" CODE WAS OBSERVED IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S BLOWER MOTOR AND SYSTEM BOARD WILL BE REPLACED TO ADDRESS THE ISSUE. DEVICE HAS BEEN EVALUATED BUT NOT REPAIRED, PENDING CUSTOMER APPROVAL OF THE ESTIMATE.

Description of Event or Problem · 1

THE MANUFACTURER RECEIVED INFO ALLEGING A VENTILATOR WOULD NOT DELIVER AIRFLOW. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
559005 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1054260

Patients

Seq Age Sex Outcome Treatment
1