FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 4110257
·
Received September 10, 2014
Report
- Report Number
- 2518422-2014-01563
- Event Type
- Malfunction
- Date Received
- September 10, 2014
- Date of Event
- August 18, 2014
- Report Date
- August 18, 2014
- Manufacturer
- RESPIRONICS INC.
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVAL OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "VENTILATOR INOPERATIVE" CODE WAS OBSERVED IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S BLOWER MOTOR AND SYSTEM BOARD WILL BE REPLACED TO ADDRESS THE ISSUE. DEVICE HAS BEEN EVALUATED BUT NOT REPAIRED, PENDING CUSTOMER APPROVAL OF THE ESTIMATE.
Description of Event or Problem · 1
THE MANUFACTURER RECEIVED INFO ALLEGING A VENTILATOR WOULD NOT DELIVER AIRFLOW. THERE WAS NO HARM OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 559005 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS INC. | 1054260 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |