FDA Adverse Event Malfunction Summary report: N

PRISM HBSAG

MDR report key: 4110165 · Received September 23, 2014

Report

Report Number
3002809144-2014-00176
Event Type
Malfunction
Date Received
September 23, 2014
Date of Event
June 25, 2014
Report Date
September 3, 2014
Manufacturer
ABBOTT GERMANY
Product Code
KSJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PO
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. AN EVALUATION IS IN-PROCESS.

Additional Manufacturer Narrative · 1

ON OCTOBER 23, 2014 IT WAS IDENTIFIED THAT: THE INCORRECT MANUFACTURING LOCATION WAS DOCUMENTED IN THIS MANUFACTURER REPORT. THE REPORT SHOULD HAVE INDICATED IT WAS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 03A47, THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 06D19, WHICH HAS A DIFFERENT MANUFACTURING LOCATION. A NEW MANUFACTURER REPORT (NUMBER 1415939-2014-00235) HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER. AN EVALUATION IS IN-PROCESS.

Description of Event or Problem · 1

THE CUSTOMER OBSERVED (B)(6) RESULTS WHILE USING THE PRISM (B)(4) ASSAY. THE CUSTOMER PROVIDED THE FOLLOWING DATA FOR THE (B)(6) YEAR OLD MALE DONOR: (B)(6) 2014: (B)(6). THIS DONOR GENERATED (B)(6) RESULTS USING THE NAT AND BIORAD ELISA METHODS. IT WAS DOCUMENTED THAT THE DONOR WAS VACCINATED FOR (B)(6) AT THE AGE OF 12. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
590336 PRISM HBSAG HBSAG KSJ ABBOTT GERMANY 37107LI00

Patients

Seq Age Sex Outcome Treatment
1 25 YR LN 06A36-04, SN (B)(4)| PRISM 6 CHANNEL ANALYZER