FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 4110158 · Received September 11, 2014

Report

Report Number
1052693-2014-00311
Event Type
Malfunction
Date Received
September 11, 2014
Date of Event
August 15, 2014
Report Date
December 21, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT NOT YET RETURNED FOR EVALUATION. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE OR USER REUSED TEST STRIP OR USER'S TEST STRIP HAD POOR FILL.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF BLOOD RESULT READING OF "LO." ANOTHER TEST PERFORMED ALSO READ "LO." NO ADVERSE EVENT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF BLOOD RESULT READING OF "LO". ANOTHER TEST PERFORMED ALSO READ "LO". NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
563278 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PP1194

Patients

Seq Age Sex Outcome Treatment
1 0 YR