FDA Adverse Event
Malfunction
Summary report: N
TRUERESULT
MDR report key: 4110158
·
Received September 11, 2014
Report
- Report Number
- 1052693-2014-00311
- Event Type
- Malfunction
- Date Received
- September 11, 2014
- Date of Event
- August 15, 2014
- Report Date
- December 21, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PRODUCT NOT YET RETURNED FOR EVALUATION. (B)(4).
Additional Manufacturer Narrative · 1
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE OR USER REUSED TEST STRIP OR USER'S TEST STRIP HAD POOR FILL.
Description of Event or Problem · 1
CONSUMER COMPLAINT OF BLOOD RESULT READING OF "LO." ANOTHER TEST PERFORMED ALSO READ "LO." NO ADVERSE EVENT REPORTED.
Description of Event or Problem · 1
CONSUMER COMPLAINT OF BLOOD RESULT READING OF "LO". ANOTHER TEST PERFORMED ALSO READ "LO". NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 563278 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PP1194 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |