TRUERESULT
Report
- Report Number
- 1052693-2014-00306
- Event Type
- Malfunction
- Date Received
- September 11, 2014
- Date of Event
- August 14, 2014
- Report Date
- December 21, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT NOT YET RETURNED FOR EVALUATION. (B)(4).
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE OR USER REUSED TEST STRIP OR USER'S TEST STRIP HAD POOR FILL. PRODUCT CODES UPDATED.
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CALLER NORMALLY IS AROUND 90-118MG/DL ONE HOUR AFTER MEALS. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE LOWEST RESULT GIVEN (60) AND THE NORMAL RESULT (118) IS LOCATED IN ZONE B/C. NO ADVERSE EVENT REPORTED.
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CALLER NORMALLY IS AROUND 90-118MG/DL ONE HOUR AFTER MEALS. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE LOWEST RESULT GIVEN (60) AND THE NORMAL RESULT (118) IS LOCATED IN ZONE B/C. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 563265 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PP1135 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |