FDA Adverse Event Malfunction Summary report: N

OSCILL-SAW-ATTACHM F/532.001 532.010 532

MDR report key: 4110139 · Received September 23, 2014

Report

Report Number
8030965-2014-01091
Event Type
Malfunction
Date Received
September 23, 2014
Report Date
March 7, 2013
Manufacturer
SYNTHES GMBH
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED A COUPLING SIDE DEFECT. THE ATTACHMENT GETS HOT. THE ATTACHMENT IS DIRTY ON THE INSIDE AND SEVERAL ELEMENTARY PARTS ARE WORN-OUT AND DIRT GETS IN. THE ATTACHMENT WAS CLEANED AND PARTS WERE REPLACED. THE DEVICE WAS RETURNED TO THE CUSTOMER ON (B)(4) 2013. PLACEHOLDER.

Description of Event or Problem · 1

DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE, TECHNICIAN INDICATED THE ATTACHMENT GOT HOT. THIS IS REPORT 1 OF 1 FOR COMPLAINT#(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
590294 OSCILL-SAW-ATTACHM F/532.001 532.010 532 HWE SYNTHES GMBH 6062

Patients

Seq Age Sex Outcome Treatment
1