FDA Adverse Event Malfunction Summary report: N

CD HORIZON SPINAL SYSTEM

MDR report key: 4110001 · Received September 23, 2014

Report

Report Number
1030489-2014-03730
Event Type
Malfunction
Date Received
September 23, 2014
Date of Event
August 22, 2014
Report Date
August 26, 2014
Manufacturer
MEDTRONIC SOFAMOR DANEK PUERTO RICO OPERATIONS
Product Code
KWP
PMA / PMN Number
K981676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): NEITHER THE DEVICE NOR APPLICABLE IMAGING FILMS WERE RETURNED TO THE MANUFACTURER FOR EVALUATION, THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFO: MACROSCOPIC AND OPTICAL EXAMINATION DID NOT IDENTIFY THREAD CREST OR FLANK DAMAGE. FUNCTIONAL EVALUATION WITH A SAMPLE MAS FOUND THE IMPLANT WAS ABLE TO BE FULLY ENGAGED INTO THE MAS HEAD WITHOUT ISSUE. AFTER VISUAL, OPTICAL AND FUNCTIONAL EVALUATION, NO EVIDENCE WAS FOUND THAT WOULD SUGGEST A DEFECT IN MANUFACTURING OR PROCESSING OF THE IMPLANT OR ASSOCIATED COMPONENTS. THE IMPLANT APPEARS TO BE CAPABLE OF PERFORMING ITS INTENDED FUNCTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A POSTERIOR SPINAL SURGICAL PROCEDURE. IT WAS REPORTED THAT THE SET SCREW SLIPPED DURING INSERTION INTO THE BONE SCREW. THE SCREW WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
590436 CD HORIZON SPINAL SYSTEM APPLIANCE, FIXATION, SPINAL INTERLAMINAL KWP MEDTRONIC SOFAMOR DANEK PUERTO RICO OPERATIONS NA H5090157

Patients

Seq Age Sex Outcome Treatment
1 00058 YR