FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4103836 · Received September 19, 2014

Report

Report Number
2032227-2014-26475
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
August 21, 2014
Report Date
August 21, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED UNABLE TO PRIME DURING PRIME/A33 TEST DUE TO MOISTURE DAMAGE ON THE FORCE SENSOR RESISTOR. UNABLE TO PERFORM EXCESSIVE NO DELIVERY TEST AND OCCLUSION TEST DUE TO PRIME FILL ANOMALY. NO MOTOR ERROR ALARM NOTED THE INSULIN PUMP HAS CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED LCD WINDOW AND CRACKED CASE AT THE DISPLAY WINDOW CORNERS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD NORMAL BLOOD GLUCOSE LEVELS OF 93.6 MG/DL. THE CUSTOMER REPORTED A MOTOR ERROR FROM THE INSULIN PUMP DURING BOLUS. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD NEED TO BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
582260 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LWWL

Patients

Seq Age Sex Outcome Treatment
1