PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-26711
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 8, 2014
- Report Date
- August 22, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND DUE TO CORRODED MOTOR HOME SWITCH. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO MOTOR ERROR ALARM. INSULIN PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE AND MISSING END CAP STICKER.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. IT WAS ALSO REPORTED THAT THE MOTOR ERROR ALARM HAD OCCURRED FREQUENTLY. AN OPERATION TEST AND THE SELF TEST WERE COMPLETED NORMALLY. NOTHING FURTHER WAS REPORTED.