PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4103744
·
Received September 19, 2014
Report
- Report Number
- 2032227-2014-25536
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 22, 2014
- Report Date
- August 22, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
INSULIN PUMP RECEIVED WITH NO ESC AND ACT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED DURING TESTING. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS AND CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
*IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE BLOOD GLUCOSE LEVEL WAS NOT INCLUDED IN THE REPORT. NOTHING FURTHER REPORTED.