FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4103660 · Received September 19, 2014

Report

Report Number
2032227-2014-25386
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
August 22, 2014
Report Date
August 22, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

ALL BUTTONS FUNCTION PROPERLY HOWEVER, MOISTURE DAMAGE WAS FOUND ON KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED DURING TESTING. INSULIN PUMP RECEIVED WITH MINOR CRACKED DISPLAY WINDOW CORNERS, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND BROKEN PUMP BELT CLIP SLOT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED FOR A BUTTON ERROR. THE BLOOD GLUCOSE READING WAS 108 MG/DL. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
581608 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-754WWH

Patients

Seq Age Sex Outcome Treatment
1