PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-25386
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 22, 2014
- Report Date
- August 22, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
ALL BUTTONS FUNCTION PROPERLY HOWEVER, MOISTURE DAMAGE WAS FOUND ON KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED DURING TESTING. INSULIN PUMP RECEIVED WITH MINOR CRACKED DISPLAY WINDOW CORNERS, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND BROKEN PUMP BELT CLIP SLOT.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED FOR A BUTTON ERROR. THE BLOOD GLUCOSE READING WAS 108 MG/DL. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 581608 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-754WWH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |