PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-25367
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 22, 2014
- Report Date
- August 22, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- PATIENT
Narratives
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. THE DEVICE WAS RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. NO TRACES OF MOISTURE WERE NOTED AT THE ELECTRONIC OR MOTOR ASSEMBLIES DURING VISUAL INSPECTION. THE UNIT WAS RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNERS, MINOR SCRATCHED LCD WINDOW AND MISSING END CAP STICKER.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED A BUTTON ERROR ALARM. SHE DID NOT RECALL THE BLOOD GLUCOSE READING. SHE STATED THAT SHE WAS IN THE SHOWER WHEN THE ALARM OCCURRED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 581768 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723LNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |