PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-25365
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 22, 2014
- Report Date
- August 22, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH NO BUTTON RESPONSE DUE TO THE CORRODED KEYPAD TRACES. NO UNEXPECTED AUDIO/BEEP OR BUTTON ERROR ALARM NOTED DURING TESTING. THE DEVICE WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND MISSING END CAP STICKER.
CUSTOMER REPORTED A BUTTON ERROR ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 116 MG/DL. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ALARM. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION IS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 582344 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-523NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |