FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4103646 · Received September 19, 2014

Report

Report Number
2032227-2014-25364
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
August 22, 2014
Report Date
August 22, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT NO BUTTON ERROR ALARMS NOTED. MINOR SCRATCHES ON THE LCD WINDOW, CRACKED CASE AT THE LCD WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, SCRATCHES ON THE RESERVOIR TUBE WINDOW, CRACKED RESERVOIR TUBE, CRACKED BATTERY TUBE THREADS, CRACKED BELT CLIP SLOT, STAINED END CAP STICKER, SHIFTED END CAP STICKER AND STAINED ADDRESS/SERIAL NUMBER LABEL. ALL THE BUTTONS DID FUNCTION PROPERLY. HOWEVER, MOISTURE DAMAGE ON THE KEYPAD TRACES NOTED.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 150 MG/DL. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ALARM. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
582623 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523NAP

Patients

Seq Age Sex Outcome Treatment
1 52 YR