FLOGARD
Report
- Report Number
- 1416980-2014-32493
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 27, 2014
- Report Date
- August 27, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION, FUNCTIONAL TESTING, BATTERY TESTING, AND A REVIEW OF THE ALARM LOG WERE PERFORMED. FUNCTIONAL TESTING AND THE REVIEW OF THE ALARM LOG REVEALED THAT THE DEVICE HAD PRESENTED AN F-94 ALARM. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE A LOW BATTERY. TO CORRECT THE CONDITION, THE MAIN BATTERY WAS REPLACED AND THE CONFIGURATION SETTINGS WERE RESET. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A FLOGARD INFUSION PUMP PRESENTED AN F-94 ALARM. IT WAS NOT SPECIFIED WHEN IN THE PROCESS THIS OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 583896 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |