FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 4103529
·
Received September 19, 2014
Report
- Report Number
- 1416980-2014-32488
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 27, 2014
- Report Date
- August 27, 2014
- Manufacturer
- SHARP CORPORATION OSAKA JAPAN
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS SERVICED AT THE CUSTOMER SITE BY A FIELD SERVICE TECHNICIAN. THE DEVICE WAS VISUALLY INSPECTED, FUNCTIONALLY TESTED AND THE ALARM LOG WAS REVIEWED. THE REPORTED CONDITION OF AN F-38 ALARM WAS VERIFIED. THE CAUSE OF THE REPORTED CONDITION WAS DUE TO DAMAGED FORCE SENSING RESISTORS (FSRS). TO CORRECT THE ISSUE THE FSRS WERE REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT A FLOGARD INFUSION PUMP EXPERIENCED AN F-38 ALARM. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 583859 | FLOGARD | PUMP, INFUSION | FRN | SHARP CORPORATION OSAKA JAPAN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|