FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 4103529 · Received September 19, 2014

Report

Report Number
1416980-2014-32488
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
August 27, 2014
Report Date
August 27, 2014
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED AT THE CUSTOMER SITE BY A FIELD SERVICE TECHNICIAN. THE DEVICE WAS VISUALLY INSPECTED, FUNCTIONALLY TESTED AND THE ALARM LOG WAS REVIEWED. THE REPORTED CONDITION OF AN F-38 ALARM WAS VERIFIED. THE CAUSE OF THE REPORTED CONDITION WAS DUE TO DAMAGED FORCE SENSING RESISTORS (FSRS). TO CORRECT THE ISSUE THE FSRS WERE REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A FLOGARD INFUSION PUMP EXPERIENCED AN F-38 ALARM. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
583859 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1