PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-26376
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 21, 2014
- Report Date
- August 21, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- PATIENT
Narratives
FAILURE ANALYSIS REVEALED THAT NO MOISTURE DAMAGE FOUND INSIDE THE INSULIN PUMP. ALL THE BUTTONS DID FUNCTION PROPERLY AND NO DAMAGE ON THE KEYPAD ASSEMBLY. THE DEVICE WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT THE DISPLAY WINDOW CORNER, AND CRACKED RESERVOIR TUBE LIP.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED THE INSULIN PUMP WAS EXPOSED TO MOISTURE. CUSTOMER STATED SHE WENT INTO A HOT TUB WEARING THE INSULIN PUMP. CUSTOMER REPORTED THERE IS FLUID UNDER THE LCD DISPLAY. BLOOD GLUCOSE VALUE IS 115 MG/DL. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 584494 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723LNAH | A000289605 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |