FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4103414 · Received September 19, 2014

Report

Report Number
2032227-2014-25277
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
August 21, 2014
Report Date
August 21, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW, SCRATCHED RESERVOIR WINDOW AND CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS. NO BUTTON RESPONSE DUE TO FLATTENED DOME SWITCH ON THE ACTIVATE BUTTON. NO BUTTON ERROR ALARMS NOTED.

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM THAT HE WAS UNABLE TO CLEAR. THE BLOOD GLUCOSE READING WAS 124 MG/DL. HE ALSO REPORTED THAT THE ACT BUTTON WAS NOT FULLY RESPONSIVE. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
584754 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 67 YR