FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4103387 · Received September 19, 2014

Report

Report Number
2032227-2014-25255
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
August 21, 2014
Report Date
August 21, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. NO TRACES OF MOISTURE WERE NOTED AT THE ELECTRONIC OR MOTOR ASSEMBLIES DURING VISUAL INSPECTION. THE DEVICE ALSO WAS RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS, AND MINOR SCRATCHED DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING A BUTTON ERROR ALARM ON THE INSULIN PUMP. CUSTOMER REPORTED THAT THE INSULIN PUMP HAD RECENTLY BEEN EXPOSED TO SWEAT. BLOOD GLUCOSE LEVEL AT THE TIME OF THE CALL WAS 111 MG/DL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
583964 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723NAP

Patients

Seq Age Sex Outcome Treatment
1 37 YR