FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 4103379 · Received September 19, 2014

Report

Report Number
3004753838-2014-29276
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
December 9, 2013
Report Date
December 9, 2013
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RETURNED COMPLAINT DEVICE WAS VISUALLY INSPECTED AND FOUND NO OBSERVATIONS RELATED TO THE COMPLAINT. A REVIEW OF THE RECEIVER DATA LOG CONFIRMED A HARDWARE ERROR CODE. THIS COMPLAINT WAS DEEMED REPORTABLE UPON COMPLETION OF DEVICE EVALUATION ON (B)(4)2014. THE ROOT CAUSE WAS DETERMINED TO BE A HARDWARE COMPONENT FAILURE.

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2013 TO REPORT HAVING TO PRESS THE CENTER BUTTON MULTIPLE TIMES ON THE RECEIVER TO GET THE SCREEN TO APPEAR ON (B)(6) 2013. PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION AT THE TIME OF CONTACT WITH DEXCOM TECHNICAL SUPPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
584354 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT20649

Patients

Seq Age Sex Outcome Treatment
1 61 YR