FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4103376 · Received September 19, 2014

Report

Report Number
2032227-2014-25251
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
August 21, 2014
Report Date
August 21, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH UNRESPONSIVE BUTTONS AND BUTTON ERROR ALARM DUE TO CORRODED KEYPAD TRACES. THE DEVICE HAD CRACKED LCD WINDOW, CRACKED DISPLAY WINDOW CORNER, CRACKED BELT CLIP SLOT AT BATTERY TUBE THREADS AREA, CRACKED RESERVOIR TUBE LIP AND CRACKED RESERVOIR TUBE.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

CUSTOMER'S HUSBAND REPORTED THAT THE INSULIN PUMP RETURNED A BUTTON ERROR ALARM. CUSTOMER DID NOT RECALL ANY SIGNIFICANT EVENTS LEADING UP TO THE ERROR ALARM. BLOOD GLUCOSE LEVEL WAS 122 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.