530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-25211
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 21, 2014
- Report Date
- August 21, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP HAD NO BUTTON RESPONSE DUE TO A FLATTENED ESCAPE BUTTON DOME SWITCH. NO WEAK SIGNAL ALARM COULD BE VERIFIED DUE TO THE UNRESPONSIVE BUTTONS. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED DISPLAY WINDOW.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED RECEIVING A WEAK SIGNAL ALARM ON THE INSULIN PUMP. CUSTOMER ALSO REPORTED THAT THE ESCAPE BUTTON WAS HARD TO PRESS. BLOOD GLUCOSE LEVEL WAS 203 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 584310 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC MINIMED | MMT-551NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR |