530G INSULIN PUMP

MDR report key: 4103357 · Received September 19, 2014

Report

Report Number
2032227-2014-25211
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
August 21, 2014
Report Date
August 21, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD NO BUTTON RESPONSE DUE TO A FLATTENED ESCAPE BUTTON DOME SWITCH. NO WEAK SIGNAL ALARM COULD BE VERIFIED DUE TO THE UNRESPONSIVE BUTTONS. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING A WEAK SIGNAL ALARM ON THE INSULIN PUMP. CUSTOMER ALSO REPORTED THAT THE ESCAPE BUTTON WAS HARD TO PRESS. BLOOD GLUCOSE LEVEL WAS 203 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
584310 530G INSULIN PUMP OZO OZO MEDTRONIC MINIMED MMT-551NAH

Patients

Seq Age Sex Outcome Treatment
1 3 YR