PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-25194
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 21, 2014
- Report Date
- August 21, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- OTHER
Narratives
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INSULIN PUMP WAS RECEIVED WITH UNRESPONSIVE BUTTONS AND BUTTON ERROR ALARM DUE TO CORRODED KEYPAD TRACES. NO MOISTURE DAMAGE FOUND ON ELECTRONIC ASSEMBLY OR MOTOR. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW AND CRACKED RESERVOIR TUBE LIP.
CUSTOMER REPORTED RECEIVING A BUTTON ERROR ALARM ON THE INSULIN PUMP AND STATED THAT THEY WERE UNABLE TO CLEAR IT. CUSTOMER STATED THAT THE INSULIN PUMP WAS AROUND HIS WAIST WHILE FISHING. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 221 MG/DL. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 584471 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR |