FDA Adverse Event
Malfunction
Summary report: N
G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 4103321
·
Received September 19, 2014
Report
- Report Number
- 3004753838-2014-29244
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- February 3, 2014
- Report Date
- February 20, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE RETURNED COMPLAINT DEVICE WAS VISUALLY INSPECTED AND FOUND NO OBSERVATIONS RELATED TO THE COMPLAINT. A REVIEW OF THE RECEIVER DATA LOG CONFIRMED A HARDWARE ERROR CODE. THIS COMPLAINT WAS DEEMED REPORTABLE UPON COMPLETION OF DEVICE EVALUATION ON 08/26/2014. THE ROOT CAUSE WAS DETERMINED TO BE A HARDWARE COMPONENT FAILURE.
Description of Event or Problem · 1
DISTRIBUTOR CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO REPORT A HARDWARE ERROR MESSAGE PATIENT EXPERIENCED ON (B)(6) 2014. DISTRIBUTOR DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION AT THE TIME OF CONTACT WITH DEXCOM TECHNICAL SUPPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 584595 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT20649-2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |