FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 4103308
·
Received September 19, 2014
Report
- Report Number
- 1416980-2014-32445
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 27, 2014
- Report Date
- August 27, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE FLOGARD INFUSION PUMP WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. A VISUAL INSPECTION, ALARM LOG REVIEW, FUNCTIONAL TESTING, AND A BATTERY TEST WERE PERFORMED. DURING THE EVALUATION IT WAS IDENTIFIED THAT THE BATTERY WAS DAMAGED. THE CAUSE OF THE DAMAGE WAS NOT DETERMINED. TO CORRECT THE CONDITION, THE MAIN BATTERY WAS REPLACED. THE PUMP WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP WAS FOUND TO HAVE A DAMAGED BATTERY. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 583605 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |