FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 4103247
·
Received September 19, 2014
Report
- Report Number
- 8020893-2014-02226
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Report Date
- September 3, 2014
- Manufacturer
- COVIDIEN
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) EVALUATED THE DEVICE, AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB), THE BACKLIGHT INVERTER PRINTED CIRCUIT BOARDS (PCBS), AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MANUFACTURING SPECIFICATIONS. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED BY A CUSTOMER IN THE USA THAT, THE 840 VENTILATOR GRAPHIC USER INTERFACE (GUI) DISPLAY WENT BLANK. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 581661 | 840 VENTILATOR | CONTINUOUS VENTILATOR | CBK | COVIDIEN | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |