FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4103186 · Received September 19, 2014

Report

Report Number
2531779-2014-26870
Event Type
Malfunction
Date Received
September 19, 2014
Report Date
August 28, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/17/2014 WITH THE FOLLOWING FINDINGS: THERE WAS NO DAMAGE OBSERVED TO THE KEYPAD. THE KEYPAD BUTTONS WERE TESTED AND WERE CONFIRMED TO BE FULLY FUNCTIONAL. THE KEYPAD WAS REMOVED AND NO BUTTON CONTACT DEFECTS WERE OBSERVED.

Additional Manufacturer Narrative · 1

THE PRODUCT HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE UP, DOWN, AND OK BUTTONS WERE UNDER RESPONSIVE TO BUTTON PRESSES. THE REPORTER INDICATED THAT THERE WAS NO KNOWN MOISTURE INGRESS TO THE PUMP AND CONFIRMED THAT THERE WAS NO ALLEGED IMPACT ON THE PUMP¿S INSULIN DELIVERY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
582435 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 24 YR