(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. THE HIGH DRAIN 101 (NIGHT DRAIN #1) EVENT WAS VERIFIED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED AS NOT ENOUGH EVIDENCE WAS AVAILABLE TO MAKE A DETERMINATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN ERROR 101 (NIGHT DRAIN #1) ALARM WAS IDENTIFIED IN THE LOG (INCLUDES ANY STANDARD MODE THERAPY WHERE THE PATIENT DRAIN VOLUME EXCEEDS 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME). THE ALARM OCCURRED ON (B)(6) 2014 AT 21:39:20. NO ADDITIONAL INFORMATION IS AVAILABLE.