ARCHITECT I1000SR ANALYZER
Report
- Report Number
- 1628664-2014-00210
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Report Date
- August 28, 2014
- Manufacturer
- ABBOTT MANUFACTURING INC
- Product Code
- JJE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. HIGH TEST RESULTS PATIENT PROBLEM; NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. THE COMPLETE SID IS: (B)(6). AN EVALUATION IS IN PROCESS.
FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, A REVIEW OF LABELING AND INSTRUMENT SERVICE. NO ADVERSE TREND WAS IDENTIFIED FOR THE CUSTOMER'S ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. A SINGLE DEFINITIVE CAUSE WAS NOT IDENTIFIED. FIELD SERVICE CONSIDERED THE CA 19-9XR RESULT ISSUE RESOLVED THROUGH REPLACING THE VORTEXER, STABILIZED (PART NUMBER 7-76656-02) WHICH WAS REPLACED DURING TROUBLESHOOTING AND WAS CONSIDERED WORN OUT FROM NORMAL USE. HOWEVER SAMPLE HANDLING COULD NOT BE RULED OUT AS A CONTRIBUTING FACTOR; ONLY 1 PATIENT RESULT WAS IMPACTED. MULTIPLE PARTS WERE REPLACED TO INCLUDE VORTEXER 1 AND WZ VALVES 1, 2 AND 3 AS A PRECAUTION. BASED ON ALL AVAILABLE INFORMATION AND ABBOTT DIAGNOSTICS' COMPLAINT INVESTIGATION, THE INSTRUMENT PERFORMED AS INTENDED AND NO PRODUCT DEFICIENCY WAS IDENTIFIED.
THE CUSTOMER OBSERVED A FALSELY ELEVATED CA19-9XR RESULT FOR ONE PATIENT ON THE ARCHITECT I1000SR ANALYZER. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(6). (NOTE: PREVIOUS VALUE: 2 U/ML); INITIAL: 55.7 U/ML, REPEAT 3.8 U/ML, 5.4 U/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 582694 | ARCHITECT I1000SR ANALYZER | AUTOMATED IMMUNOASSAY ANALYZER | JJE | ABBOTT MANUFACTURING INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCHITECT CA19-9 REAGENTS| LIST 02K91-25 LOT 33394M500| ARCHITECT CA19-9 REAGENTS| LIST 02K91-25 LOT 33394M500 |