FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4103101 · Received September 19, 2014

Report

Report Number
2032227-2014-26412
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
May 3, 2014
Report Date
August 26, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4). THE INSULIN PUMP WAS RECEIVED WITH OPERATING CURRENTS WITHIN SPECIFICATION. THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/A33 TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. NO UNEXPECTED LOW BATTERY ALARM OR EXCESSIVE NO DELIVERY ALARMS WERE NOTED. THE INSULIN PUMP WAS RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS. IT WAS RECEIVED WITH A MINOR SCRATCHED DISPLAY WINDOW AND MISSING END CAP STICKER.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE HAD RECEIVED NO DELIVERY ALARMS AND THAT THE BATTERIES ONLY LASTED ABOUT TWO DAYS. THE BLOOD GLUCOSE READING WAS 135 MG/DL. NOTHING FURTHER WAS REPORTED.