FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4103080 · Received September 19, 2014

Report

Report Number
2032227-2014-26410
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
April 22, 2014
Report Date
August 26, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INSULIN PUMP FAILED THE DISPLACEMENT TEST (170.6 UNITS LEFT ON STATUS SCREEN) AND MOTOR ERROR ALARMS DURING REWIND DUE TO OUT OF PHASE MOTOR ENCODER SIGNAL. NO A35 ALARMS NOTED. CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON DISPLAY WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED.

Description of Event or Problem · 1

CUSTOMER'S HUSBAND REPORTED THAT THE CUSTOMER HAD A MRI AND THE INSULIN PUMP WAS NOT REMOVED. HE STATED THAT THE INSULIN PUMP ALARMED FOR A MOTOR ERROR. THE BLOOD GLUCOSE READING WAS 104 MG/DL. HE WAS ASSISTED IN CLEARING THE ALARM. HE STATED THAT THE DRIVE SUPPORT CAP WAS NORMAL. HE STATED THAT THEY WERE NOT ABLE TO REWIND THE INSULIN PUMP. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
582359 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522NAB

Patients

Seq Age Sex Outcome Treatment
1 42 YR