CORE IMPACTION DRILL
Report
- Report Number
- 0001811755-2014-03304
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- August 11, 2014
- Report Date
- August 25, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K112593
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE TECHNICIAN WAS ABLE TO DUPLICATE THE REPORTED EVENT OF OVERHEATING AND NOTED THAT THE DRIVESHAFT BEARINGS WERE DAMAGED, WHICH MAY HAVE CONTRIBUTED TO THE REPORTED EVENT.
THE DEVICE IS AVAILABLE FOR EVALUATION; ADDITIONAL INFORMATION MAY BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETE. THE DEVICE IS AVAILABLE FOR EVALUATION; ADDITIONAL INFORMATION MAY BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETE.
IT WAS REPORTED THAT DURING A CASE AT THE USER FACILITY, THE CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICALLY SIGNIFICANT DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
IT WAS REPORTED THAT DURING A CASE AT THE USER FACILITY, THE CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICALLY SIGNIFICANT DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 582628 | CORE IMPACTION DRILL | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |