FDA Adverse Event
Summary report: N
840 VENTILATOR
MDR report key: 4102975
·
Received September 19, 2014
Report
- Report Number
- 8020893-2014-02165
- Date Received
- September 19, 2014
- Date of Event
- January 1, 2014
- Report Date
- August 15, 2014
- Manufacturer
- COVIDIEN
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
COVIDIEN RECEIVED INFO STATING THAT AN 840 VENTILATOR EXPERIENCED AN ERRATIC LOWER DISPLAY. THE VENTILATOR WAS NOT IN USE ON A PT AT THE TIME THE MALFUNCTION OCCURRED.