FDA Adverse Event Malfunction Summary report: N

SUMMIT UNIVERSAL BROACH HANDLE

MDR report key: 4102808 · Received September 19, 2014

Report

Report Number
1818910-2014-28523
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
September 16, 2014
Report Date
September 16, 2014
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE RETURNED DEVICE WAS UNABLE TO CONFIRM THE REPORTED EVENT. THE RETURNED HANDLE WAS FUNCTIONALLY TESTED WITH A DEPUY RETAINED BROACH AND ASSEMBLED/DISASSEMBLE AND HELD THE BROACH FIRMLY. THE OVERALL CONDITION OF THE HANDLE SUGGESTS HEAVY USAGE. A COMPLAINT DATABASE IDENTIFIED SIMILAR REPORTS AGAINST THE PROVIDED PRODUCT CODE WHICH WERE ATTRIBUTED TO PRODUCT WEAR OUT AND OR MISUSE. THE ROOT CAUSE IS ATTRIBUTED TO HEAVY USAGE/WEAR OUT DUE TO THE OVERALL CONDITION OF THE RETURNED DEVICE AND THE LENGTH OF TIME IN SERVICE. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE BROACH HANDLE DOES NOT STAY LOCKED WHEN ATTACHED TO BROACH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
584381 SUMMIT UNIVERSAL BROACH HANDLE HIP INSTRUMENT LXH DEPUY ORTHOPAEDICS, INC. 1818910 AK0502

Patients

Seq Age Sex Outcome Treatment
1